Dansensor
MAP Mix Provectus
®
Location Register read
0xA07A
M1 position
ADC10 value
(Read only)
0xA07C
M2 position
ADC10 value
(Read only)
0xA005
Voltage input
range:
2 - 10V or 0 - 10V
(Read only)
DE
Encoding
2 bytes unsigned integer
Range 0 to 1023
This is the external voltage input signal controlling the
setting of the CO
mixing for a 3-gas mixer.
2
For a CO
/O
mixer it controls CO
2
2
For N
/O
and CO
/O
mixers it controls O
2
2
2
2
Input range is 0 to 11.75V ~ 0 to 1023 ADC counts
For input range 2 to 10V for 0 to 100% mix, the ADC read-out
range will be approx. 174 to 871 counts
For input range 0 to 10V for 0 to 100% mix, the ADC read-out
range will be approx. 0 to 871 counts
Input must change more than 50mV from current setting
before new mix settings is recalculated.
2 bytes unsigned integer
Range 0 to 1023
Only present in a 3-gas mixer.
This is the external voltage input signal controlling the
setting of the O
mixing for a 3-gas mixer.
2
Input range is 0 to 11.75V ~ 0 to 1023 ADC counts
For input range 2 to 10V for 0 to 100% mix, the ADC read-out
range will be approx. 174 to 871 counts.
For input range 0 to 10V for 0 to 100% mix, the ADC read-out
range will be approx. 0 to 871 counts.
Input must change more than 50mV from current setting
before new mix settings is recalculated.
1 byte unsigned.
Readout of the range used for the two voltage input signals
controlling the mixer setting.
0 = range is 0-10V
1 = range is 2-10V (default)
Bedienungsanleitung
(gas 1).
2
(gas 2).
2
P/N 300863-H
12/2019
39