5. Commissioning, operation

5. Commissioning, operation
Personnel: Skilled personnel
Tools: Slotted screwdriver or open-ended spanner
WARNING!
Physical injuries and damage to property and the
environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene,
flammable or toxic substances), harmful media (e.g. corro-
sive, toxic, carcinogenic, radioactive), and also with refrige-
ration plants and compressors, there is a danger of physical
injuries and damage to property and the environment.
Should a failure occur, aggressive media with extremely high
temperature and under high pressure or vacuum may be
present at the instrument.
For these media, in addition to all standard regulations,
the appropriate existing codes or regulations must also be
followed.
5.1 Mounting instructions for instruments with EHEDG and 3-A
Observe the following instructions, especially for EHEDG-certified and
3-A-compliant instruments.
Mount the instrument with minimal dead space and able to be
cleaned easily.
The mounting position of the instrument, welding socket and instru-
mentation T-piece should be designed to be self-draining.
The mounting position must not form a draining point or cause a basin
to be formed.
Mounting position with slight downward inclination so that the
cleaning agents wet all parts and run off cleanly.
With the process connection via an instrumentation T-piece, the
branch L of the T-piece must not be longer than the diameter D of the
T-piece (L ≤ D).
WIKA operating instructions model TG58SA
EN
13