Herunterladen Inhalt Inhalt Diese Seite drucken

HEIDENHAIN EIB 5000 Betriebsanleitung Seite 59

Vorschau ausblenden Andere Handbücher für EIB 5000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 6
7
What to do if ...
Fault conditions
The product provides a comprehensive monitoring function. Fault conditions are indicated through changes in the initial
temperature value. During the initialization phase after power-on, the maximum value is output for the temperature.
After this phase, the temperature value levels out at the actual measured value.
EIB 521x
Temperature: ≤ 200 °C = valid temperature values
Temperature : ≥ 220 °C = general fault
Reaction
Temp
= -158.1 °C
max
Temp
= 253.x °C
max
Temp
= 254.0 °C
max
Note: PTC/PTC triplet: no detection possible because terminating resistance is also high-impedance
1)
HEIDENHAIN | EIB 5000 | Operating Instructions | 04/2021
Fault condition
Fault
≤ 30.0
Sensor short
R
Sensor
≥ 6500.0
R
Sensor wire break-
1)
Sensor
age
x= 0: sensor 0
x= 1: sensor 1
x= 2: sensor 2
Invalid configuration
Corrective action
Check the sensor and sensor connection;
exchange, if necessary
Check the sensor and sensor connection;
exchange, if necessary
Contact HEIDENHAIN
59

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis