Troubleshooting
Replace product: have defective products repaired by Aesculap Technical Service, see Technical Service.
Fault
Battery not working
Battery quickly empty

10. Storage and transport

WARNING
Risk of injury from burns and damage to product in case of short-cir-
cuit!
When storing the product, ensure that the product cannot short-
circuit with other products or metal objects.
In order to prevent damage during transport, ship product in the orig-
inal packaging and in an outer carton.
If the original packaging is no longer available, ensure the following
when packaging the product:
– The battery contacts are protected against short-circuit.
– The packaging completely encloses the product.
– The product is secured against movement within the packaging.
Li-Ion batteries must be identified with corresponding warning notices on
the shipment packaging when sending through the mail.
Identify product with warning notice when shipping:
– UN3480, if only batteries are in the shipment packaging.
– UN3481, if machines and batteries are in the shipment packaging.
Due to the regulations of the International Air Transport Association
(IATA), this product may not be returned by air shipment.
Ship product only by ground/sea transport.
Never ship defective batteries.
Dispose of defective products, see Disposal.
Note
Batteries with normal wear (decreased capacity) that have no evidence of
damage are not considered defective. These batteries can be shipped for a
service case to Aesculap Technical Service.
Due to transport regulations, the product is charged only approx. 30 % and
has a limited storage life of 12 months. After the first complete charge, the
product can be stored for 24 months without recharging.
Completely charge the product upon receipt and before extended peri-
ods of storage.

11. Accessories/Replacement Parts

Art. no.
Designation
TA014536
Instructions for use, Acculan 4 Li-Ion battery short
GA346 (A4)
TA014537
Instructions for use, Acculan 4 Li-ion battery short
GA346 (leaflet)
6
Detection
No signal tone when inserting the bat-
tery into the machine
Symbol "Battery replacement recom-
mended" illuminates at the end of the
charging process on the charger GA320
Cause
Battery broken
Battery cells at end of life

12. Technical Service

DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
Do not modify the product.
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.

13. Technical Data

13.1 Classification according to Regulation (EU)
2017/745
Art. no.
Designation
GA346
Acculan 4 Li-Ion battery short
13.2 Performance data, information and standards
Model
Acculan 4 Li-ion battery short GA346
Cell type
Li-ion
DC voltage
9.9 V
Energy content
10,89 Wh
Weight
approx. 0.265 kg
Dimensions (L x W x
approx. 120 mm x 43 mm x 50 mm
H)
EMC
IEC/DIN 60601-1-2
Conforming to
IEC/DIN 60601-1
standard
Remedy
Replace product.
Have battery cells replaced by
Aesculap Technical Service.
Class
I