4.

Product description

4.1
Scope of Supply
Art. no.
Name
GA346
Acculan 4 Li-Ion battery short
TA014537
Instructions for use Acculan 4 Li-ion batteries
4.2
Components required to operate the device
Acculan 4 charger GA320
Acculan 4 or Acculan 3Ti machine:
– GA340
– GA341
– GA344
– GT666
– GA670
– GA671
– GT668
– GA675
4.3
Operating principle
The Acculan 4 Li-Ion battery supplies the Acculan machines with short
handle with voltage. The battery is charged in the Acculan 4 charger
GA320. The charging process of a completely discharged battery can take
up to 120 min.
The speed and rotational direction are set depending on the trigger setting
of the machines:
for saws and Dermatome: speed clockwise
for drills: speed clockwise, counter-clockwise, oscillation and pilgrim
step
5.
Preparation
Prior to use, inspect the product and its accessories for any visible dam-
age.
Use the products and their accessories only if they are in perfect tech-
nical condition.
6.

Working with the device

6.1
System set-up
See instructions for use for the respective Acculan machine and the
Acculan 4 charger GA320.
6.2
Functional test
See the instructions for use for the respective Acculan machine.
6.3
Safe operation
WARNING
Risk of injury and material damage due to inappropriate use of bat-
teries!
Charge battery only in the Acculan 4 charger GA320.
Insert battery correctly (Acculan machine and charger).
WARNING
Risk of injury and material damage due to damaged batteries!
Check batteries for any damage before charging.
Do not use or recharge damaged batteries.
4
CAUTION
Total battery failure can occur if the batteries are left in the charger
while it is disconnected from the power supply!
Always keep the charger connected to the power supply. Indicator
light for readiness for service must be lit.
Always remove batteries from the charger when it is disconnected
from the power supply.
CAUTION
Loss of capacity/performance due to prolonged storage of discharged
batteries!
Leave the batteries in the functional charger after they have been
successfully charged (trickle charging).
For proper loading and removal of the battery from the Acculan
machine, see the instructions for use for the respective Acculan
machine.
7.

Validated reprocessing procedure

7.1
General safety information
Note
Observe national statutory regulations, national and international stan-
dards and directives, and any proprietary hygiene instructions for process-
ing.
Note
The applicable national regulations concerning the reprocessing of devices
must be observed for patients having Creutzfeldt-Jakob disease (CJD), sus-
pected of having CJD, or possible variants of CJD.
Note
Please note that successful processing of this medical device can only be
guaranteed after the processing method has been validated. The opera-
tor/processor shall be responsible.
Note
If there is no final sterilization, then a virucidal disinfectant must be used.
Note
See also B. Braun eIFU at eifu.bbraun.com for current information on pro-
cessing and compatibility of material
7.2
General notes
Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product's
materials according to the chemical manufacturers' recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
Optical changes of materials, e.g. fading or discoloration of titanium or
aluminum. For aluminum, the application/process solution only needs
to be of pH >8 to cause visible surface changes.
Material damage such as corrosion, cracks, fracturing, premature aging
or swelling.
Do not use metal cleaning brushes or other abrasives that would dam-
age the product surface and could cause corrosion.