9. Specifications

9.2 Overload safety
Model
PG43SA-S 2 x full scale value, max. 40 bar or max. pressure rating
EN
PG43SA-C 5 x full scale value, max. 40 bar or max. pressure rating
1) see PN in the tables of data sheets PM 04.15, PM 04.16
9.3 Vacuum resistance (only model PG43SA-S)
All scale ranges are suitable for the vacuums typical of CIP or SIP
cleaning. Positive scale ranges < 6 bar [< 100 psi] are optionally
available in a vacuum-resistant version (to -1 bar).
9.4 Permissible temperature
Place, situation
Ambient
Medium
CIP and SIP
Autoclaving
Storage, transport
1) only for autoclavable version
9.5 Temperature effect
Model
PG43SA-S
PG43SA-C
18
Overload safety
(PN) of the process connection
(PN) of the process connection
Permissible temperature
-20 ... +60 °C
≤ 150 °C
150 C continuously for wetted parts
≤ 134 °C, ≤ 20 minutes
1)
-40 ... +70 °C
Deviation from the reference temperature
(+20 °C) at the measuring system
≤ ±0.5 %/10 K of full scale value
≤ ±0.8 %/10 K of full scale value