Herunterladen Inhalt Inhalt Diese Seite drucken

Device Parameters And Information - thyracont VSM77E Betriebsanleitung

Vakuum transmitter
Vorschau ausblenden Andere Handbücher für VSM77E:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

37

5.4 Device Parameters and Information

Type of Device (TD):
Query of device type, e.g. VSM207
Product Name (PN):
Query of product name (article number)
Serial Number Device (SD):
Query of device serial number
Serial Number Head (SH):
Query of sensor head serial number
Version Device (VD):
Query of the device's hardware version
Version Firmware (VF):
Query of the device's firmware version
Version Bootloader (VB):
Query of the device's bootloader version
Baud Rate (BR):
Set the baud rate for data transmission
Value range: 9600, 14400, 19200, 28800, 38400, 57600, 115200 Bd
Response Delay (RD):
Query and set the time delay between receiving a telegram and sending the ans-
wer.
Value range: 1 ... 99999 µs (default 5500 µs)
Device Restart (DR):
Make a device reset
) Not accessible via EtherCAT
VSM-EtherCAT-200116

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Vsm78eVsm79e

Inhaltsverzeichnis