Enabling the Option 001 SOP
1) In measuring mode, press the Menu softkey.
2) Select "Configuration" and confirm by pressing enter.
3) Select Option "001 SOP" and enter your activation code.
Adapt SOP Cal
Here, you specify which buffers are to be used in which sequence.
You can combine buffer solutions from different buffer sets. Please note that the
minimum distance allowed between two buffer solutions is Δ 2 pH.
The "Configuration - Calibration" menu is extended as follows:
+ Calibration
Cal mode
Adapt SOP cal
Cal points
Buffer 1
Buffer set
Buffer
Buffer 2
Buffer 3
Verification
Delta pH
Verification buffer
Calimatic | Manual | Data entry | SOP cal
1-point | 2-point | 3-point
Mettler-Toledo
Knick CaliMat
Ciba
NIST technical
NIST standard
Hach
WTW
Hamilton
Reagecon
DIN 19267
Metrohm
User buffer 1
Select a buffer from the selected set
Select buffer set 2 and buffer (see buffer 1)
Select buffer set 3 and buffer (see buffer 1)
Off | On
pH 0.05
(Enter maximum permitted deviation from
verification buffer; exceeding this value
generates an error message)
Select buffer set and buffer (see buffer 1)
2.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01/12.00
2.00/4.01/7.00/10.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
4.00/7.00/9.00
45