13. EMC / Technical description
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
162
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
3 A/m
Mains power qual-
<5 % U
ity should be that of a
T
(>95 % dip in U
)
typical commercial or
T
for 0,5 cycle
hospital environment.
40 % U
If the user of the elec-
T
(60 % dip in U
)
tric breastpump Swing
T
for 5 cycles
requires continued
70 % U
operation during power
T
(30 % dip in U
mains interruptions, it
)
T
is recommended that
for 25 cycles
<5 % U
the electric breastpump
T
(>95 % dip in U
)
Swing is powered from
T
for 5 s
an uninterruptible power
supply or a battery.
Power frequency mag-
netic fields should be
at levels characteristic
30 A/m
of a typical location in
a typical commercial or
hospital environment.