Operating Matrix - Endress+Hauser Levelflex FMP 232 Schnellinbetriebnahme

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

2. Calibration for internals with a partially filled silo,
so that a calibration can also be carried out if the probe is partially covered and
has been shortened.
1.) Go to field V9H5
• Enter "333" while keeping the
+
simultaneously press
I
G
2.) Go to field V3H0 (3x
press)
I
• Enter "2" (partial calibration if the probe is free)
• Press
key
G
3.) Enter "1" in field V3H1
4.) Go to field V3H5 (4x
press)
G
• Enter free length of probe in m, distance to material min. 0.5 m
+
press simultaneously, the Setup is now evaluated
I
G
• Wait until the bar graph no longer flashes,
approx. 30 - 60 s
5.) Go to field V3H5
• Enter the entire length of the probe – 0.35 m as lower range-value
6.) Go to field V0H1
• Enter the same value as in V3H5 (zero adjustement)
• Press
key
G
7.) Enter measuring range (span) in field V0H2 , (min. distance to process connection: 300 mm, check value in field V3H8)

3. Operating matrix

H0
H1
V0
Measured value
Basic
calibration
Technical
unit
[%]
(26)
[V3H5]
V1
V2
Linearisation
0:Height
Linaris-
1:Activate
2:Table
ation
3:Semi-automatic
4:Clear
5:Linear
[5]
(27)
V3
Calibration
Mode
Extended
0: Factory calibra-
calibration
tion
1: Customer
2: Partial Calibration
[0]
(24)
[0]
V4...V6
V7
*
Service
V8
Operating
0: 4...20 mA 1: 4 mA
Mode
limit 2: 4/20 mA 3: 8/
[0]
V9
Diagnostic
code
diagnostic code
Simulation
(32)
VA
Tag No.
Remote
operation
V99
System
The Levelflex is now calibrated.
Checking measurement reliability:
The strength of the signal can be checked after the silo has been filled. Go to field V3H2. If a value >= 3 is shown here then measurement is assured over
a long period of time. With very shortened probes, the value for the blocking distance should be checked in field V3H8 and compared with the measuring
range to ensure that the blocking distance lies outside the measuring range.
Additional information: "http://www.endress.com"
Endress+Hauser
key pressed down, (reset to default values)
O
H2
H3
Empty
Full calibration
calibration
(26)
[0,9 x E]
(26)
Line No.
Enter level
1-11
Technical
meter/feet
(27)
(27)
Calibration
Reflection
Actual level
quality
0...10
0: Ready
1: Activate
(24)
(24)
[E=m]
*
Assign current
Select distance
Delay time
output mode
units:
0: m
1: ft
for E 641
16 mA
(28)
[0]
(26)
[100]
Last
Device and
software
(32)
Units after
linearisation
1...12:%, l, hl
3
m
qft (=ft
ft, US-gal
H4
H5
Output damping
Value for 4 mA
(8mA)
Seconds
0-250
Technical
unit
[5]
(28)
[0]
Enter
Max.-volume
volume
Technical
unit
unit
(27)
Measuring lenght
(range)
m/ft
E
1...10 m
1...20 m
(-)
not used
*
*
*
s
(28)
Device
Reset
address
333: customer
version
(30)
(-)
Material
, dm, cm
3
), kg, t
H6
H7
Value for 20 mA
Safety Alarm
(16 mA)
0:MIN
1:MAX
2:HOLD
Technical
unit
(28)
[100]
(28)
[2]
(28)
(26)
Probe type
Process connection
0: uncoated
0: standard
1: coated
(24)
(-)
(-)
*
*
Simulation
Simulation
0: off
value
1: Level
2: Volume
3: current
(35)
[0]
(34)
(34)
FMP 232, 332
H8
H9
Measured distance
Level
(D) meter/feet
(L1)
meter/feet
(6)
(6)
Blocking distance
Persistence filter
Meter/Feet
(35)
[5]
(34)
*
*
Current output
Security lock
333: unlock
mA
xxx: lock
(34)
(22)
2

Werbung

loading

Diese Anleitung auch für:

Levelflex fmp 332