Levelflex - Quick Setup
The Levelflex is delivered with a factory calibration so that, in general, the instrument operates without requiring further
calibration. A calibration / modification to the application is required if:
1.) The probe is mounted in a nozzle which is > 100 mm high and / or ∅ > 100 mm
2.) The probe is shortened
3.) Distance from probe to internals > 400 mm
1. Calibration for internals with an empty silo
1.1 Without display or display removed,
probe not shortened !
1.) Reset to
default values:
2.) Calibration
1.2 With display and probe shortened
Entries into "matrix fields",
Move around the matrix with
I
Entering values with
or
keys
O
S
Activate function so that the display flashes,
by pressing the
and
keys!
O
S
1.) Go to field V9H5
• Enter "333" while keeping the
down, (reset to default values)
+
press simultaneously
I
G
2.) Go to field V3H0 (3x
press)
I
• Enter "1" (complete calibration for the whole
length of the probe)
• Press
key
G
3.) Enter "1" in field V3H1
4.) Go to field V3H5 (4x
press)
G
• Enter length of probe in m, otherwise value already entered:
Press
or
key.
O
S
+
press simultaneously, the Setup is
I
G
now evaluated
• Wait until the bar graph no longer flashes,
approx. 30 - 60 s
If the length of the probe has been changed:
5.) Go to field V3H5
• Read off value for lower range-value
6.) Go to field V0H1
• Enter value from V3H5 (zero adjustment)
• Press
key
G
7.) Enter measuring range (span) in m in field V0H2,
(min. distance to process connection: 300 mm) Check value in field V3H8
1
press till red LED lights up,
wait till red LED goes out
(approx. 30-60 s)
press till red LED lights up,
wait till red LED goes out
(approx. 30-60 s)
and
Keys
G
key pressed
O
L: Probe lenght ordered
B: Blocking distance, min.: 30 cm (V3H8) F: Full distance = span (V0H2)
Measured value
Bar graph
(current output/visible monitoring)
L+
L+
L-
L-
L1
L1
N
N
1
1
2
2
+
V
FMP 232, 332
E: Empty distance = zero (V3H5)
Matrix position
V V H H
4 ...
4 ...
20 mA
20 mA
I+
I+
I-
I-
4
4
5
5
6
!
Red LED
Fault
GreenLED
Standby
H
+