3-3. Connection Cable Installation
Use a tightening torque of 1.0 N·m for the connector that is used for the connection with the sensor head.
Note
If an additional tightness of cable connections is required, it is possible to use a thread locking adhesives
to the threaded area of the connector, i.e. Loctite 242 or its equivalent.
Note
When the connector has been connected with a torque of 1.0 N·m, check to see that the O-ring of the
connector is correctly placed inside the sensor head as indicated below.
SQ47-Z Series
14
(E)
Fig. 3-13
Sensor head
Sensor head
Connector
Fig. 3-14
Tightening torque: 1.0 N·m
1
2
Connector
Sensor head
Connector
O-ring
SQ47-Z Series, CH22 Series