6. Apply a thread-locking agent to the threaded areas, and install the sensor head so that the cable
extension direction matches that confi rmed in step 4.
Notes
Take care that the reading part of the sensor head does not interfere with the scale.
At this point it is recommended to insert a 0.185 mm clearance gauge, loosely turn the mounting
screws, and check the clearance between the scale and the reading part of the sensor head.
<When using the stop surfaces>
Sensor head bracket
Sensor head
Scale bracket
7.
Use the supplied clearance gauge and check that the clearance between the scale and the reading part
of the sensor head is 0.185
* Three types of clearance gauges (0.1 mm, 0.185 mm, and 0.25 mm) are supplied. After adjusting the clearance with the
0.185 mm gauge, make sure that the 0.1 mm gauge can be inserted and the 0.25 mm cannot.
<When using the stop surfaces>
Scale bracket
Scale
SQ47-Z Series, CH22 Series
Sensor head bracket
M6 × 25
(9 N m)
M6 × 25
Thread-locking
(9 N m)
agent
Fig. 3-8
+ 0.065
mm.
– 0.085
Clearance gauge
Sensor head
Fig. 3-9
<When using parallel pins>
M6 × 25
(9 N m)
Sensor head
Scale bracket
<When using parallel pins>
Scale bracket
M6 × 25
(9 N m)
Thread-locking
agent
Unit: mm
Clearance gauge
Unit: mm