60 minutes

7. CLEANING AND DISINFECTION

9. Therapy has been completed.
(You will hear 3x beep-sound).
Turn device off by pressing the On/Off button for longer
than 3 seconds.
10. Carefully remove the strap (COMSfix/COMSrefix)and
COMS One from patient and dispose the single-use
material (COMStouch & COMSfix).
1 1. Clean and disinfect device before using it on next patient
(see next chapter).
12. Let the device rest before the next use (60min). This
prevents heating of the device at high ambient tempe-
ratures and allows the disinfectant to act optimally.
CAUTION
• Do not apply device more than once within 12 hours on
the same location to prevent overstimulation of the tissue.
WARNING
• Unplug the device before cleaning and disinfection.
• Do not clean, disinfect or perform other service and
maintenance tasks during use of the device.
The COMS One device is reusable. Thorough cleaning
and disinfection between use on different patients is
important.
1.
Cleaning: Wipe off with a clean, damp cloth (water or
non-abrasive detergent).
ENGLISH
13