fabricate a prosthetic limb,
the following aspects have
to be considered:
• Use components only
according to their intended
purpose.
• If components with different
weight limits are combined,
the weight limit of the
weakest component applies
to the whole prosthesis.
• If components for different
activity levels are combined,
the activity level of the
component with the lowest
activity level applies to the
whole prosthesis.
• Using individual, certified
components with the
CE-marking does not release
technicians from their duty
to check component combi-
nations within their means
for fitness for purpose,
proper assembly and safety.
• Should there be evidence
that a certain combination
of components is not as safe
EN-4
as required, the components
User Manual
• The prosthesis has to be
• The locking torque specifica-
• Before the prothesis is handed
• Safety related regulations for
• If different components (from
must not be combined.
configured and assembled
according to the recognized
technical rules of the trade.
tions given for the modular
component have to be met.
A suitable tool (i.e. a torque
wrench) has to be used to set
the locking torque accordingly.
over to the user, all modu-
lar components have to be
secured against accidental
loosening or twisting with
suitable Threadlocking fluid.
individual components (e.g.
specific combination possibi-
lities, maintenance intervals
etc.) must be followed.
one or from several manu-
facturers) are combined: Only
use components which meet
the requirements of DIN EN
ISO 10328 and the Medical
Devices Act.