WARTUNG
4.5
Kalibrierung
The calibration of the control sensor is traceable to national standards. Calibrated
replacement control sensors can be ordered from Michell Instruments, or the system can
be adjusted with respect to any other hygrometer by means of the internal calibration
correction routine (see section 3.4.5.4).
A fi ve-point calibration certifi cate for the internal HygroSmart HS3 control sensor is
provided as standard with each instrument. In addition to this, the system is tested
as a whole at three points against a reference chilled mirror hygrometer. It is possible
to request the HygroCal100 with UKAS certifi cation, contact your Michell Instruments
representative for further information.
Certificate of Conformity
Certificate of Conformity No.
Product Description
Product Order Code
Product Model Serial No.
Sales Order No.
Line No.
Configured Probe Functional Test Completed
Calibration Certificate
Interchangeable Sensor Order
Interchangeable Sensor Serial
Interchangeable Sensor Cal Date
The above mentioned item has been calibrated at the following points in the Michell Instruments Humidity Calibration Laboratory against
the below Test Equipment traceable to the defined National Standard.
Calibration Reference Equipment
The uncertainties are based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of confidence of
approximately 95%.
Michell Instruments certify that the above equipment has been designed, manufactured, tested and inspected in full
accordance with our quality assurance procedures and conforms to the requirements of the contract/purchase order.
Name
36
Reference Humidity %RH
Observed Humidity %RH
15.10
30.10
50.15
70.20
90.44
Reference Temperature °C
Observed Temperature °C
23.39
Type
Humidity Generator
Humidity Reference
Electrical Reference
Temperature Reference
Frank Lauwerijssen
Michell Instruments Benelux B.V. Krombraak 11, Oosterhout, 4906 CR, The Netherlands
Abb 36
Hygrocal100 Bedienungsanleitung
10000019
Humidity Calibrator Portable
HC100+B1+C0+D1+E3+F1+F5+F7+F8
PAA000321
22303
1
15 March 2016
HS3-S
SAA000338
15 March 2016
Difference %RH
15.56
0.46
30.79
0.69
50.85
0.70
70.77
0.57
90.55
0.11
Difference °C
23.45
0.06
Serial Number
Calibration Date
GENHR01
11/07/12
ETALTD02
01/08/13
MNUM5708
05/02/15
ETALTS03
20/11/14
Quality Technician
www.michell.com
Kalibrierzertifi kat (Beispiel)
Permissible Difference %RH
+0.8/-0.8
+0.8/-0.8
+0.8/-0.8
+0.8/-0.8
+0.8/-0.8
Permissible Difference °C
+0.2/-0.2
Certificate No.
2379MBW2012
H1318910E
C07E150295
TCB001
97500 Ausgabe 3, Juli 2018