6. Pull the red tape to remove the seal.
7.  Pull off the sterile paper. Do not touch the spring
at this stage.
8. Hold the MiniMed Mio 30 with one hand and pull
the lid off with the other.
CAUTION: Be careful not to bend or touch the
MiniMed Mio 30 introducer needle. If the infusion
set is not securely placed in the insertion device
with the needle pointing straight ahead, pain or
minor injury may occur during insertion.
9. Place your fingers on the finger grips on each
side. With your other hand, use your finger to
gently pull back the spring until you hear a "click".
The needle guard will lift up or peel off. Do not
touch or bend the introducer needle. Check that
the soft cannula does not exceed the introducer
needle.
10. Be sure the adhesive tape is not stuck on the
introducer needle.
11. INSERTION: Place your index finger on the top
release button and your thumb on the bottom
release button. Position MiniMed Mio 30 with its
legs resting flat on the skin to ensure an insertion
angle of 30°. Make sure to maintain a 30° angle
while inserting and gently press the release
buttons once to insert MiniMed Mio 30.
12. Make sure that the cannula stays in place. Gently
use two fingers to hold down the tape while
removing the insertion device and introducer
needle. Gently pull the insertion device straight
back.
13. Secure the cannula by gently pressing down
with one finger on the clear window of the tape.
Carefully remove the protective liner under the
cannula housing.
14. Remove the backing paper under the front end.
Gently massage the adhesive tape onto the skin.
15. Put the lid back onto the MiniMed Mio 30 for
disposal.
16. Gently place a finger on the cannula housing
to hold it and insert the site connector into the
infusion set using your other hand. Make sure
the connector locks with a "click". Fill the empty
space in the cannula with 0.7 units (0.007 ml) of 
U-100 insulin.
Disconnecting
MiniMed Mio 30 allows you to temporarily disconnect
your pump without having to change the infusion set.
A. Gently place a finger in front of the cannula
housing and squeeze both sides of the site
connector. Pull the connector straight out.
B. Insert the protective cap onto the site connector
until you hear a "click".
C. Insert the disconnect cover into the cannula
housing until you hear a "click".
Reconnecting
D. Remove the disconnect cover from the cannula
housing.
E. Remove the protective cap from the site
connector. If necessary, fill the infusion set until
insulin dispenses from the needle.
CAUTION: When filling, hold MiniMed Mio 30
with the needle pointing down.
F. Gently place a finger in front of the cannula
housing. Push the site connector into the cannula
housing until you hear a "click".