Cautions
Observe the following Cautions to avoid product damage that
could otherwise lead to safety concerns:
• Use only compatible accessories described in this Instructions
for Use (IFU).
• Do not immerse or sterilize the M2501A Sensor, or any
Airway Adapter, except as directed in this IFU.
• Before using a reusable Airway Adapter, check that the
Adapter windows are dry and residue free, and have not been
damaged from handling, cleaning, or sterilizing.
• Do not apply any tension to the M2501A Sensor cable.
• Do not exceed storage or operating temperature limits.
• Observe precautions for electrostatic discharge (ESD) and
electromagnetic interference (EMI) to and from other
equipment.

Troubleshooting Tips

Check the following whenever a change in patient ETCO
waveform is observed:
• Check patient status.
• Check positioning of Endotracheal Tube.
• Ensure Sensor is correctly connected to Airway Adapter.
• Check that Airway Adapter is not damaged and is securely
connected to ventilator circuit.
• If problem is not identified, change the Airway Adapter.
• If problem is still not identified, change the M2501A Sensor.
3