CO
Sensor Adapter Zero
2
An "Adapter Zero" is a quick procedure that adjusts the
M2501A Sensor to the optical characteristics of each compatible
Airway Adapter. The procedure is necessary only:
• When switching the M2501A Sensor from one Airway
Adapter type to another, such as when switching from a
disposable to a reusable Airway Adapter.
• When requested by the CO
For additional information about the Adapter Zero procedure,
reference the CO
monitor's Instructions for Use.
2
monitor.
2
Maintenance and Testing
Refer to the CO
monitor's Installation and Service Guide for
2
maintenance and testing of the M2501A Sensor.
NOTE: Test procedures in the CO
specify use of the gas and gas cylinder regulator listed below.
Gas Cylinder Regulator and Verification Gas
M2506A Gas
Gas used to check M2501A Sensor
operation.
M2505A Gas
Provides gas flow control when using
Cylinder Regulator
M2506A Gas to check M2501A Sensor
operation.
7