4-4-1. Detailed Description of MODE Switches

MODE switches 1, 2, 4, 7, 9 and 10
These switches are not used. Always use in the OFF setting.
MODE switch 3 (Direction switch)
This switch is used to change the positional relationship between scale movement direction and A/B signal.
For more information, see section 4-5-1, "Changing the Direction."
MODE switch 5
This switch is not used. Always use in the ON setting.
MODE switch 6 (Resolution setting switch)
This switch is used to set the resolution. For output form F types, setting this switch to ON sets 0.1 µm and
setting to OFF sets 0.05 µm. For G types, setting to ON sets 0.02 µm and setting to OFF sets 0.01 µm.
MODE switch 8 (Alarm mode switch)
This switch is used to select the alarm output format.
For more information, see section 4-5-3, "Setting the Alarm Reset Mode."

4-5. Changing the Settings (For output format F, and G only)

4-5-1. Changing the Direction

MODE switch 3 (Direction switch) is used to change the positional relationship of the A/B signal with
respect to the scale movement direction.
MODE switch 3
OFF
ON
For output format A, refer to section 4-6-1.

4-5-2. Changing the Resolution

MODE switch 6 (resolution setting switch) is used to change the resolution of the A/B signal output type.
In the case of output format F: ON = 0.1 µm
4-4
(E)
A
B
A
B
Table 4-2
OFF = 0.05 µm
Moves to head cable side
A signal is leading
B signal is leading
In the case of output format G: ON = 0.02 µm
OFF = 0.01 µm
BL57-NE