n C160 : 1000 A calibre (2000 A max. peak)
Intensity in A~
Intrinsic error
Dephasing
C160 : 100 A calibre (300 A max. peak)
Intensity in A~
Intrinsic error
Dephasing
C160 : 10 A calibre (30 A max. peak)
Intensity in A~
Intrinsic error
Dephasing
(1) As % of output signal
n C173 : 1000 A calibre
Intensity in A~
Intrinsic error
Dephasing
C173 : 100 A calibre
Intensity in A~
Intrinsic error
Dephasing
C173 : 10 A calibre
Intensity in A~
Intrinsic error
Dephasing
C173 : 1 A calibre
Intensity in A~
Intrinsic error
Dephasing
(1) As % of output signal
4.4

Conditions of use

The "C100" clamps must be used in the following conditions, in order to comply with user safety and
metrological performance requirements.
1...50 A
(1)
(2)
0.1...5 A
(1)
(2)
0.1...0.5 A
(1)
(2)
1...10 A
£ 1% + 0.2 mV
(1)
(2)
0.1...1 A
£ 1% + 0.2 mV
(1)
(2)
0.01...0.1 A
£ 1% + 0.2 mV
(1)
(2)
0.001...0.01 A
£ 3% + 1 mV
(1)
(2)
50...200 A
£ 1% + 1 mV
£ 3°
5...20 A
£ 2% + 5 mV
£ 15°
0.5...2 A
£ 3% + 10 mV
(2)
(2) Unspecified
10...100 A
£ 0,5% + 0.2 mV
£ 2°
1...10 A
£ 0.5% + 0.2 mV
£ 2°
0.1...1 A
£ 0.5% + 0.2 mV
£ 5°
0.01...0.1 A
£ 3% + 1 mV
(2)
(2) Unspecified
13
200...1000 A
1000...1200 A
£ 2°
20...100 A
£ 10°
2...10 A
£ 15°
100...1000 A
1000...1200 A
£ 0.2%
£ 1°
10...100 A
£ 0.3%
£ 1°
1...10 A
£ 0.5%
£ 2°
0.1...1 A
£ 0.7% + 1 mV
£ 0.7% + 1 mV
£ 10°
£ 1°
100...120 A
£ 5°
10...12 A
£ 15°
£ 0.2%
£ 1°
100...120 A
£ 0.2%
£ 1°
10...12 A
£ 0.5%
£ 2°
1...1.2 A
£ 10°