Ultrasonic cleaning is strictly prohibited.
2. Disinfection
Soaking in solutions or Thermo chemically in a washer Sterilizer up to 65°C maximum may perform disinfection.
Manufacturers instruction regarding duration and concentration of solutions should be strictly adhered. After Disin-
fection, rinse throughly in sterile water and dry with a lint free clean cloth.
3. Cold Soak Solution
To achieve a high-level disinfection, Cidex®OPA or 2.4% Glutaraldehyde solution may be used according to manu-
facturer's instructions. Dry with lint free, clean cloth or filtered pressurized air. Reassemble all parts, load handle with
batteries and test the system for proper function.
If not functional, review battery / lamp testing instructions below.
Do not immerse Blades in Bleach, Betadine or Potassium Hydroxide solutions.
Doing so will several damage instruments also avoid metal to metal contact after soaking, the blades should
be rinsed under sterile water to remove chemical residues and dry with lint free clean cloth or filtered pres-
surized air.
4. Sterilization
Before performing any of the procedures described below, the blade should be cleaned as described in the cleaning
Procedure.
5. Gas Sterilization
Gas sterilization by Ethylene oxide up to a maximum temperature of 65°C and 8 p.s.i, may be performed, which is
preferred especially if sterilization is to be performed regularly.
6. Steam Sterilization
Steam Sterilization can also be performed. Insert device in appropriate autoclave pouch.
(A) GRAVITY DISPLACEMENT STEAM
Temperature
121°C (250°F)
Cycle Time
30 Minutes
Dry Time
15 Minutes
Note: Do not exceed temperature of 135°C and pressure of 28 p.s.i
Flash autoclaving and hot air sterilization should be avoided as these processes will damage the instrument.
Steris Amsco V-Pro
Laryngoscope Blades and Handles are compatible with:
Amsco V-Pro 1 Low Temperature Sterilization System
Amsco V-Pro 1 Plus Low Temperature Sterilization System
Amsco V-Pro 1 Pro Max Temperature Sterilization System
Sterrad
Larynogoscope Blades and Handles are compatible with:
Sterrad 100nx System (Standard and Express Cycle)
Sterrad nx System (Standard Cycle)
Sterrad 1005 and 200 System (Short Cycle Outside US)
Sterrad 50 System
HANDLES
Cleaning/Sterilization:
Battery Handle withstand the same cold soak solution and autoclave ranges outlined in the blade section. However,
the batteries and lamp must be removed prior to disinfection/sterilization.
Battery Handles can withstand exposure to the ethlene oxide. Lamp may be cleaned with cotton ball dampened in
alcohol (IPA). The main handle and a cap will also withstand and exposure to ethylene oxide.
ENGLISH
(B) PRE-VACUUM STEAM
134°C (273°F)
5 Minutes
20 Minutes