Herunterladen Inhalt Inhalt Diese Seite drucken

Ottobock 13E200 Gebrauchsanweisung Seite 19

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
CAUTION
Distance to HF communication devices is too small (e.g.
mobile phones, Bluetooth devices, WiFi devices)
Injury due to unexpected behaviour of the product caused by
interference with internal data communication.
► Therefore, keeping the following minimum distances to these
HF communication devices is recommended:
• Mobile phone GSM 850/GSM 900: 0.99 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.7 m
• DECT cordless phones incl. base station: 0.35 m
• WiFi (routers, access points,...): 0.22 m
• Bluetooth devices (third-party products not approved by
Ottobock): 0.22 m
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning
agents.
► Only clean the product with a damp cloth and mild soap (e.g.
453H10=1 Ottobock DermaClean).
► Do not bring the surface of the electrodes into contact with a
special cleaner for Ottobock cosmetic gloves, e.g 640F12.
CAUTION
Damaged electrode cable
Injury due to unexpected product behaviour as a result of uncon­
trolled activation of the prosthesis component.
► Power down the system and contact your orthopaedic techni­
cian.
CAUTION
Use of a damaged product
Injury due to loss of product functionality.
Downloaded from
www.Manualslib.com
manuals search engine
19

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis