how you feel, or if you think the results are not accurate.
you have dropped or think you may have damaged the
9
9
device.
To perform the control solution test, do the following:
Step 1: Insert the test strip to turn on the meter.
Insert the test strip into the meter. Wait for the meter to
display the test strip and blood drop symbols.
Step 2: Press C to mark this test as a control solution test.
If you press C again, the "QC" will disappear and this test is no
longer a control solution test.
Step 3: Apply control solution.
Shake the control solution vial thoroughly before use.
Squeeze out a drop and wipe it off, then squeeze out another
drop and place it on the tip of the vial cap. Hold the meter
to move the absorbent hole of the test strip to touch the
drop. Once the confirmation window fills completely, the
meter will begin counting down. To avoid contaminating the
control solution do not directly apply control solution onto
a strip.
Step 4: Read and compare the result.
After counting down to 0, the control solution test result will
appear on the display. Compare this result with the range
printed on the test strip vial and it should fall within this
range. If not, please read the instructions again and repeat
the control solution test.