3. Insert the liner with the pads into the Pads Sentry.
4. Plug the pads cable connector into the connector
port on the FR3. Press firmly to ensure that the
connector is fully inserted in the port.
NOTE: Do not plug the pads connector into the
FR3 unless you are using the Pads Sentry. The
pads must be installed in the Pads Sentry while
connected, or the FR3 self-test will fail.
5. Tuck the coiled pads cable into the space provided.
6. If you have a spare set of pads, store the unopened
package in the spare pads compartment in the
upper lid of your FR3 system case.