B4 Troubleshooting

B4 Troubleshooting
PROBLEM
PROBLEM
The unit does not come on. The unit's power on / off
The unit does not come on. The unit's power on / off
status light and display are off.
status light and display are off.
CAUSE / SOLUTION
CAUSE / SOLUTION
The main power supply to the unit is off or the unit is
The main power supply to the unit is off or the unit is
not switched on.
not switched on.
i.
i.
If the LED on the PSU is not on it is not receiv-
If the LED on the PSU is not on it is not receiv-
ing power from the mains. Switch the mains
ing power from the mains. Switch the mains
power on.
power on.
ii. If the LED on the PSU is on switch the unit on.
ii. If the LED on the PSU is on switch the unit on.
iii. If status light and display still do not come on
iii. If status light and display still do not come on
check the unit's membrane control panel and
check the unit's membrane control panel and
cabling. Replace if faulty.
cabling. Replace if faulty.
PROBLEM
PROBLEM
Animation displaying either Unit connection or
Animation displaying either Unit connection or
Dental imaging software (fl ashing yellow) appear
Dental imaging software (fl ashing yellow) appear
on the unit display and the unit will not work.
on the unit display and the unit will not work.
CAUSE / SOLUTION
CAUSE / SOLUTION
Defective RJ45 cable. Replace.
Defective RJ45 cable. Replace.
Imaging plate unit
Imaging plate unit
B - 15
B - 15
D506724