TV 301 CONTROLLER INSTALLATION

The controller can be mounted in two position:
bottom mounting (as per the complete system)
side mounting.
To install the controller execute the following pro-
cedures.
Bottom mounting
See the following figure.
1. Turn the pump upside-down;
2. Screw the 4 fixing studs (provided with the ac-
cessories bag) in the holes on the bottom of
pump;
3. Place the controller on the studs aligning the 4
rubber holes on the top of the controller case;
4. Press the controller towards the pump;
5. Turn the pump again;
6. Plug the line cord and connect the pump cable
to the pump;
7. Connect the mating connector with the jumper
on the interlock signal to start the pump.
MOUNTING 1
Side mounting
The L-shaped bracket (P/N 969-9970) is available
as an option.
See the following figure.
1. Screw the 4 fixing studs (provided with the ac-
cessories bag) in the holes on the L-shaped
bracket;
2. Place the controller on the studs aligning the 4
rubber holes on the top of the controller;
3. Press the controller towards the bracket;
4. Turn the pump upside-down;
5. Place the bracket on the pump bottom;
6. Screw a little bit the three screws M4 in the
holes on the pump bottom;
7. Rotate the bracket until the controller is on the
chosen position;
8. Tighten the screws properly;
9. Turn the pump again;
10. Plug the line cord and connect the pump cable
to the pump;
11. Connect the mating connector with the jumper
on the interlock signal to start the pump.