Checking and resetting the hours of duty

Select "Duty hours" and click on
. The
following parameters can be selected
UV Surface
UV Recirculator
UV Supply Air
HEPA Filter
Carbon Filter
Select one parameter and press on
to see
the time of duty. A progress bar will indicate when
the device needs to be replaced.
We recommend to change the UV tubes after
1000 – 1500 operating hours as the UV intensity
decreases significantly after that time.
We recommend to change the HEPA Filter after
1500 operating hours (based on the default setting
of 0.5 m/sec of the air flow. With less or more
power, the runtime is extended or shortened, this
will be considered automatically by the runtime
clock), or earlier to avoid clogging.
We recommend to change the Carbon Filter after
3000 h or once a year.
To reset the Duty hours of the selected parameter,
press on
the progress bar and hours starts
flashing.
Press the
or
buttons to change the value to
zero. Press on
to store. Press
to go
back to the Duty hours selection parameters.
Reaching the maximum of the recommended
operating hours, the UV tubes or filters must be
replaced. Those items will be temporary indicated
on the main screen tagged by a sandglass (duty
hours exceeded). When "duty hours exceeded"
message is displayed, press
to reset the
duty hours of the items directly (short cut).
After replacing filters, UV or white light
tubes it is necessary to reset the hours of
duty in the software manually!
19