We recommend you change the HEPA filter after 1500 hours of operation. The Carbon filter should
be repaced after 3000 h of operation (based on the default setting of 0.5 m/sec of the air flow. With
less or more power, the runtime is extended or shortened, this will be considered automatically by the
runtime clock).

Replacing the Pre-filter

Turn the unit power off and unplug the unit from the power
outlet. Locate and remove the four screws on the pre-filter
cover on the left side of the workstation.
Pull away the pre-filter and replace the new pre-filter back into
position and reattach the four screws.

Replacing the fuse

The fuse is located at the back-right side of the system. To remove
the fuse, use a flathead screwdriver to open the lid and pull out the
fuse.
Insert the new fuse and using the screwdriver, turn the fuse clockwise
to lock.
26
VWR Collection PCR Workstation HEPA