The unit is factory set as follows:
MANUAL operating mode
TSP source type
sublimation current 30 A
sublimation time 1 minute
sublimation period 3 minutes
-7
pressure threshold 1x10
ìAUTOSTARTî (automatic restart before power failure) on
ìYESî
Detailed information about the operating modes are contained in
the appendix ìTechnical Informationî.
Sublimation Controller Switching ON
To switch ON the Sublimation Controller switch the power switch
located on the rear panel to the ìIî position.
To start the sublimation, push the ìSUBLIMATIONî key located
on the front panel and the relevant green LED will light.
NOTE
If the ìSUBLIMATIONî key is not set in ON position the
sublimation will not starts and the unit will wait for a parameters
setting.
Sublimation Controller Switching OFF
To switch OFF the Sublimation Controller , proceed as follows:
set the ìSUBLIMATIONî key in OFF position (green LED
extinguished)
switch the power switch located on the rear panel to the ìOî
position.
MESSAGE
POWER AC FAIL
HEATER DEFECTIVE
FILAMENT X
DEFECTIVE
CARTRIDGE
DEFECTIVE
XXX
FAULT:
OVERTEMPERATURE
EXTERNAL
INTERLOCK
mbar
DESCRIPTION
Anomalous operation relevant to the power
supply due to a frequency absence or a
frequency different from those expected.
Heater filament interruption during a pumping
cycle with the Mini Ti-Ball.
Filament X (where X can be 1, 2 or 3)
interruption during a sublimation with the TSP.
If the ìTSP RECOVERî parameter is set in
ìAUTOMATICî the software restart the normal
operation, selecting a successive cartridge
filament.
If the ìTSP RECOVERî parameter is set in
ìMANUALî the sublimation will turn off.
All TSP filament are interrupted and the
sublimation is turned off.
Transformer overload occurs and the pumping
cycle is stopped.
The interlock on pin 3 of the I/O connector is
not connected.
MAINTENANCE
The Sublimation Controller does not require any maintenance.
Any work performed on the pump must be carried out by
authorized personnel.
! ! ! !
Before carrying out any work on the controller, disconnect it from
the supply.
In the case of breakdown, contact your local Varian service
center who can supply a replacement controller.
Before returning the pump for repairs, or replacement with a
reconditioned unit, the "Health and Safety" sheet attached to this
instruction manual must be filled-in and sent to the local sales
office. A copy of the sheet must also be inserted in the pump
package before shipping.
If a controller is to be scrapped, it must be disposed of in
accordance with the specific national standards.

ERROR MESSAGES

For a certain type of failure, the controller will self-diagnose the
error and the messages described in the following table will be
displayed.
Switch OFF the unit then check the power
supply and the correct setting of voltage input
values (110 - 120 Vac, 60 Hz o 220 - 240 Vac,
50 Hz).
Replace the Mini Ti-Ball.
If the ìTSP RECOVERî parameter is set in
ìAUTOMATICî,
automatically one of the remaining filament,
and the relevant LED will be ON.
If the ìTSP RECOVERî parameter is set in
ìMANUALî the user must manually restart the
sublimation procedure.
Replace the cartridge.
Eliminates the cause of the overload on the
transformer.
Connect the interlock.
16
INSTRUCTION FOR USE
WARNING!
NOTE
REPAIR ACTION
the
software
selects
87-900-077-01(B)