Cautions
1
Federal law (U.S.A.) restricts this device to sale by, or on the
order of, a physician.
2
Stop the pump before closing the inflow cannula, closing the inflow
outlet pinch clamp (
result in damage to the pump or cassette.
3
Do not overtighten luer locks (
threads and affect fluid seal.
Setup/Connections
Warning: Always follow sterile protocol when handling sterile connections.
Inflow Cassette Tubing (blue; single-patient use only)
0450-000-100
1
Align the blue side of the cassette (
2
Insert the cassette. Push with your thumb until it clicks into place.
3
Close the spike pinch clamps (
4
Discard the protective caps from the spikes (
5
Discard the protective cap from the blue luer (
6
Open the spike pinch clamps (
7
Refer to the CrossFlow Integrated Arthroscopy Pump instructions for
use for priming, operation, and calibration instructions.
8
Close the pinch clamp (
9
Connect the blue luer (
10 Open the pinch clamp (
Outflow Cassette Tubing (red; single-patient use only)
0450-000-200
1
Align the red side of the cassette (
2
Insert the cassette. Push with your thumb until it clicks into place.
3
If using a cannula for outflow, discard the protective cap from the tube
marked
4
If using a shaver handpiece for outflow, connect the suction fitting of
the tube marked
5
If using a suction RF probe for outflow, connect the suction fitting of
the tube marked
Warning: Ensure outflow tubing is connected to the appropriate
device. The pump will not operate properly when devices are not
connected to appropriate tubing.
6
Route the waste line (
7
Close pinch clamps on any unused tubes(
Removal
1
Stop the pump.
2
Close all pinch clamps.
3
Disconnect tubing from hardware/devices.
4
Eject Inflow and/or Outflow Cassette(s).
5
Discard Cassettes and Tubing appropriately.
Sterilization
Provided sterile, by
Ethylene Oxide
Do not resterilize
Symbols
Latex free
Type BF applied part
Use by
Contains or presence
of phthalate
2
), or ejecting cassettes. Failure to do so may
6
,
5
).
3
) for bags that are spiked.
3
) near the blue luer (
6
) to the inflow cannula.
5
).
6
, and connect the red luer (
(
) to the shaver handpiece.
10
(
) to the RF probe.
11
) to a waste receptacle or collection device.
12
). Overtightening may strip the
9
) with the blue button on the pump.
1
), then spike saline bags.
4
).
5
).
5
) with the red button on the pump.
7
) to the outflow cannula.
9
).
8
Single use only
Do not use if packaging is
damaged
Catalog number
Batch code
Consult instructions for
use