Fit the valve unit onto the holder, the lid must be
open.
Fill the cleaning container with a solution of luke-
warm tap water and a commercially available denture
cleanser tablet.
Immerse the membrane and the valve (attached to
the holder) for 10 to 15 minutes (read instructions of
the denture cleanser). The container may be shaken to
facilitate cleaning.
Rinse the valve, membrane and container carefully
under tap water.
Shake off the water.
Chemical Disinfection
Fill the container with fresh 70% Ethanol, minimum
temperature 18° C / 65° F.
Immerse the valve (with open lid) and membrane.
Shake the container for a short time and let all soak for
a minimum of 30 minutes.
Tip out the Ethanol and let the valve, membrane and
container air-dry.
Storage
Put valve and membrane into the cleaning container and
leave that closed until the next use.
Instructions.for.the.Patient
See also the illustrations for patients at the front of the
manual [P1 through P18].
This patient section of the Instructions for Use de-
scribes the more regular (every-day) use of the Provox
FreeHands HME
Product Description, Indications, Contraindications,
Warnings and Adverse Events, Device Lifetime Expec-
tation, please review the general information section at
the beginning of this manual.
Please make sure that you are fully instructed and
that your clinician has adjusted the device, prior to
starting to use the Provox
any problems possibly related to the use of the Provox
FreeHands HME
STEP.1.–.PREPARATION.OF.THE.STOMA
Carefully clean and dry the skin around the stoma
using Provox
soap. Make sure that the skin is dry before proceeding
[picture P1]
Apply Skin-Prep
12
P A TIE NT S
. For all general information, e.g.,
®
®
, please consult your clinician.
®
Cleaning Towel or water and a mild
®
to the skin around the stoma
FreeHands HME
®
. In case of
®
®