PALAMIX
Feed rod
Feed rod release mechanism
Trigger grip
Description
The PALAMIX
cement gun is intended to accommodate
®
the mixing cartridges of the PALAMIX
system or the PALACOS
 R+G pro and then to ensure
®
precise and effortless application of bone cement.
The trigger of the PALAMIX
hand to expel the bone cement from the mixing cartridge.
Activating the trigger advances the feed rod in the mixing
cartridge and expels the bone cement under high pressure.
The PALAMIX
cement gun can only be used in
®
conjunction with the PALAMIX
PALACOS
 R+G pro. The respective operating steps are
®
also described in detail as follows.
Cleaning and sterilisation
The PALAMIX
cement gun is intended for multiple use
®
and must be cleaned and sterilised each time prior to use.
The feed rod can be separated from the grip part of the
PALAMIX
cement gun for ideal cleaning and sterilisation.
®
1
2
cement gun
®
vacuum mixing
®
cement gun is operated by
®
vacuum mixing system or
®
To release, unlock the
catch on the feed rod
( ) and remove the rod.
The PALAMIX
cement gun should be cleaned
®
and sterilised as for any conventional surgical instrument.
Firstly clean off gross soiling from the product using a soft
non-metallic brush. Loosen organic residues with ultrasonic
treatment and then rinse under running water. To sterilise
the PALAMIX
cement gun in the steam autoclave, sterilise
®
according to the following processing guidelines.
If the PALAMIX
cement gun has been sterilised with heat
®
or steam, it must cool down before reuse.
Maintenance
Periodic maintenance of the cement gun is not required.
Notes
The PALAMIX
cement gun should only be used by trained
®
personnel. The risk of hand injury or damage to protective
gloves is possible due to improper handling. There is also a
risk of damage to respective mixing cartridge if personel
are not trained correctly.
Check the functionality of the PALAMIX
to each operation. If damaged, contact your local Heraeus
Medical representative. The user must ensure the correct
use of the PALAMIX
cement gun.
®
4
Attachment for mixing cartridge
cement gun prior
®