NUMBERS OF RUN
[ 01 ] RUN
^
CLEAR
ENTER
Select the desired number of sampling cycles by pressing the "UP" or "DOWN" arrows.
Press the ENTER button to go to the next digit.
The maximum numbers of cycle is 19.
Press ENTER to confirm.
SINGLE RUN VOLUME
[  0000   ] Litres
^
CLEAR
ENTER
Select the desired volume of aspirated air for each sampling by pressing the "UP" or "DOWN" arrows.
Press the ENTER button to go to the next digit.
The maximum aspirated volume of the cycle is 1999 litres.
Press ENTER to confirm.
The display will show the TOTAL volume of air to be aspirated at the end of the cycles.
Check if it is the correct volume.
If not, press the CLEAR button and start again.
For example if you set 5 cycles of 200 litres each, you will aspirate a total volume of 1000 litres of air.
TOTAL RUN
[    1000    ]    Litres 
CLEAR
ENTER
Pressing ENTER to confirm will set the SAS in multi mode.