Herunterladen Inhalt Inhalt Diese Seite drucken

Automatic User-Initiated Test; Placing And Securing The Fr3; Directions For Use - Philips HeartStart FR3 Installations-, Gebrauchs- Und Wartungsanleitung

Vorschau ausblenden Andere Handbücher für HeartStart FR3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 47

AUTOMATIC USER-INITIATED TEST

Press ON for Patient Use
The first time you set up the FR3, the FR3 automatically runs a detailed user-initiated test as soon as
the battery is in place. This test is designed to make sure the device and all connected accessories are
Press Shock Button
to Run User-Initiated Test
ready for use. The first part of the test is interactive. Follow the prompts to press the buttons as
instructed, or the test will fail. Then listen for two audio tones. If you do not hear the tones, there is a
problem with the speaker or the beeper. Contact Philips at www.philips.com/AEDSupport for
technical support. The second part of the test is automatic. The FR3 displays a progress bar indicating
User-Initiated Test
the test's progress.
CHECKING SYSTEM
IMPORTANT NOTE: Do not remove the battery during the user-initiated test. To abort the test,
press the On/Off button to turn off the device.
NOTE: Do not connect or disconnect any accessories, such as the pads or Infant/Child Key, during
the user-initiated test.
When the test is complete, the FR3 displays the results of the test, the device serial number, and the
software revision, then displays the ADMINISTRATION screen. If for any reason the test fails, contact
Philips at www.philips.com/AEDSupport for technical support.
Press the On/Off button to display the status screen and put the defibrillator into standby. If the FR3
passes the test, the green Ready light flashes to show the FR3 is ready for use. If the FR3 or any
User-Initiated Test Results
connected accessory fails the test for any reason, the screen displays information about the failure.
Test Passed
Ready for Use

PLACING AND SECURING THE FR3

Serial Number
CYYM-#####
Place the FR3 in an accessible area where the green Ready light is easily visible and the FR3 alert chirps
Software Version
X##.##
can be heard. Philips recommends that the defibrillator and pads be secured in a protective carry case,
Administration
OFF
such as the FR3 rigid system case, soft system case, or small soft case.
Press ON for Patient Use
IMPORTANT NOTE: To ensure that the FR3 is ready for use when needed, store it with the battery
Shutting Down
installed and the SMART Pads III connected and installed in the Pads Sentry. Do not remove the battery
unless the FR3 self-tests indicate that it should be replaced.
Battery
Good
Data Card
Good

DIRECTIONS FOR USE

Administration
OFF
To use the FR3 in an emergency:
1. TURN ON THE FR3. Press the green On/Off button.
NOTE: If you store the defibrillator in a HeartStart FR3 case, the FR3 automatically turns on when
you open the case lid. If for any reason the FR3 does not turn on automatically, or if you do not use
an FR3 case, press the green On/Off button to turn on the FR3.
2.
FOLLOW THE VOICE AND TEXT PROMPTS. The FR3 will direct you in each step of using it to
treat a suspected victim of SCA. Place the pads as illustrated on page 5 of this manual. If the victim
is under 55 lbs (25 kg) or 8 years old, insert the Infant/Child Key, if available, in the FR3 before
placing the pads.
9

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis