Invia
Motion
®
EN
The Invia Motion Quick Card does not replace the Invia Motion instruc-
tions for use. For information concerning intended use, indications,
contraindications, warnings and cautions please refer to the appropriate
Invia Motion Clinician instructions for use and Invia Motion Patient in-
structions for use included with the Invia Motion pump or at www.medela.
com/healthcare.
Overview
Clinician
use only
Positioning and wearing Invia Motion Negative Pressure Wound Therapy
(NPWT) pump
P
P
Upright
Laying with display
side up
Shoulder strap
If placed in an inappropriate position or when canister is full, "Canister full" notification will appear
and Invia Motion canister/tubing set must be changed to continue the therapy.
Quick Card
™
Clinician and
patient use
O
Upside down
Laying with display
side down
Belt strap
Charge battery for the first time before starting the therapy
(approximate charging time is 4 hours).
Recharge battery throughout the length of the therapy. The expected
battery time generally exceeds 10 hours.
Information on the display
Symbol
Description
Pump is running/On-state in intermittent mode (Symbol is blinking)
Constant mode
Negative pressure setting in mmHg
mmHg
Battery charge level (symbol steady)
Battery charging (bars moving)
Off-state in intermittent mode (Symbol is blinking)
O
Intermittent mode
Battery fully charged
Info used time – shows runtime already used in days (
(Example for pump that has operated for 6 days and 23 hours).
Not applicable for Invia Motion Endure.
5
) and hours (
)