4.4 Find optimal feeding rate
The test mode allows you to find the optimal feeding rate and therefore to ensure, that only one part at
a time is fed to the balance. Proceed as follows to adjust the feeding rate using the test mode:
§
The device is in state "ready" (LED Ready is lit up).
1 To enter the test mode, press and hold the button Feed
Rate/Test for at least 5 seconds.
ð The LV12 starts delivery until a part passes through the
light barrier. It then stops for 2.5 seconds and then
continues delivery until the next passage through the
light barrier.
2 By briefly pressing the button Feed Rate/Test, you can now
select the rate so that only one part at a time falls onto the
balance.
3 To quit the test mode, press the button Start/Stop/
Discharge.
4.5 Configure feeding rate sets
During the configuration, the intensity level is indicated by the four green LEDs as follows:
To configure the feeding rate set, proceed as follows:
§
The device is in state "ready" (LED Ready is lit up).
1 Press and hold the two buttons for at least 5 seconds.
ð The LED Ready goes out.
2 Change the feeding rate set by toggling using the button
Feed Rate/Test.
3 To quit the setting mode, press anew the two buttons.
ð The LED Ready is lit up.
Automatic Feeder
General (for most parts, factory
settings)
Midsize parts
Small parts
Large parts
Operation 17