Appendix C
E M C I n f o r m a t i o n
The DEXIS Titanium sensor is, like any electronic medical device, subject
to electromagnetic interactions with other electronic devices. The informa-
tion in this chapter addresses this issue.
The EMC information in this chapter is provided for the medical system es-
tablished by connecting the DEXIS Titanium sensor to a computer. This
computer must be compliant with IEC 60950-1 (if located outside the pa-
tient environment) or IEC 60601-1 (if located inside the patient environ-
ment). Please consult the documentation of the computer for completing
the EMC information.
The DEXIS Titanium sensor is suitable for use in hospitals except for near
active HF surgical equipment and the RF shielded room of an ME system for
magnetic resonance imaging, where the intensity of EM disturbance is
high.
Important:
Portable/mobile radio frequency communications equipment
can affect the function of the DEXIS Titanium sensor as well as any other
electronic medical equipment. This affect may result in Image failure or im-
age quality degradation.
DEXIS is a USB compliant device and shall be used with USB compliant ca-
bles suitable for high speed/USB 2.0 cables. Such cables are either
marked "USB 2.0" or "USB high speed." USB certified hubs can be used to
extend the distance to the USB host/computer. The length of the cable
connection to the hub or between hubs shall not exceed 5 m.
WARNING
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could
result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in
improper operation.
English
35