5. Commissioning, operation
Mounting is only permitted in the depressurised state. Reliably
depressurise the instrument using the available valves and protective
devices (e.g. shut-off valve, monoflange).
Nominal position per EN 837-3 / 9.6.6 figure 7: 90° ( ⊥ ). Tolerance: ±5°
With parallel male threads (e.g. G ½ B), use flat gaskets, lens-type
sealing rings or WIKA profile sealings.
With tapered threads (e.g. NPT threads), sealing is made in the
threads, using a suitable sealing material (EN 837-2).
When screwing the instruments in, the force required for sealing must
not be applied through the case, but only through the spanner flats
provided for this purpose, and using a suitable tool.
Sealing face
The tightening torque is dependent on the selected process
connection.
WIKA operating instructions pressure gauge, model 2, copper alloy
Spanner flats
Sealing
in the thread
Installation with
open-ended spanner