Herunterladen Diese Seite drucken

Sirona inLab MC XL Handbuch Seite 15

Vorschau ausblenden Andere Handbücher für inLab MC XL:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Sirona Dental Systems GmbH
Using the cleaning set CEREC MC XL, inLab MC XL
61 77 161 D 3439
D 3439.103.01.07.02 02.2010
1.
Unscrew the chuck and remove the milling instrument.
2.
Clean the inside and outside of the chuck with a supplied pointed cotton
swab and undiluted ethanol (commercially available cleaning alcohol).
When doing this, rotate the cotton swab counterclockwise in order to
push the dirt out.
3.
Clean the thread with a clean cloth and undiluted ethanol (commercially
available cleaning alcohol).
4.
Grease the thread very lightly.
5.
Clean the inside of the clamping cone with a pointed cotton swab and un -
diluted ethanol (commercially available cleaning alcohol). When doing
this, rotate the cotton swab counterclockwise in order to push the dirt out.
6.
Insert the milling instrument in the chuck and screw in the chuck by hand.
Screw the milling instrument tight with the torque wrench.
7.
Repeat these steps until you have cleaned all of the milling instrument
chucks.
Using the cleaning set
Milling instrument chucks
5