Using the cleaning set CEREC MC XL, inLab MC XL
61 77 161 D 3439
D 3439.103.01.07.02 02.2010
1
Using the cleaning set
1.1
Cleaning the manual block clamp
➢ Clean the inside of the block axle once a month with a round (green) cot -
ton swab and undiluted ethanol (commercially available cleaning alco -
hol).
1.2
Milling instruments without chuck
CAUTION
Do not grease the thread.
Do not grease the thread when using milling instruments without chucks.
NOTICE
For CEREC MC XL: Once a month
For inLab MC XL: Once a week or after every 4th water change
Clean the clamping cones of the milling instruments.
The motors are in the position for changing the milling instruments (burs).
1.
Unscrew and remove the milling instrument.
2.
Clean the inside of the clamping cone with a pointed cotton swab and un -
diluted ethanol (commercially available cleaning alcohol). When doing
this, rotate the cotton swab counterclockwise in order to push the dirt out.
3.
Do not grease the thread. Insert the milling instrument and screw it tight
with the torque wrench.