Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

CELL PROGRAMMER

The 7 Cell Programmer is fitted as standard and enables up to seven samples to
be presented for measurement sequentially.
This section describes the operation of the accessory with the software and also
describes its removal / refit.
When the Cell Programmer is installed then a status box appears on the top right
hand side of the screen, to the left of the Absorbance/Wavelength box, on all the
pages (except HOME) indicating the the presence of the accessory and its setting.
If this status box is displayed, then the Cell Programmer can be moved manually
using the < > arrow keys. To step foward/backward many positions, repeated
pressing of the key will be 'remembered' by the system.
* CELL
PROG *
CELL POS.
: 1
MODE
: MANUAL
REF. MODE
: OFF
LAST CELL
: 7
CELL CYCLES
: 5
SPEED
: HIGH
CELL POS.
MODE
MANUAL
RUN&STEP
AUTO
OFF
The number of cells automatically measured in sequence (maximum of 7)
will be dependent on the value of the LAST CELL parameter
REF. MODE
When ON, in all the above modes CELL 1 is assigned as the REFERENCE position
and a zero or baseline will be performed.
LAST CELL
CELL CYCLES
42
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
CELL PROGRAMMER
This page allows the Cell Programmer to be
set up for the required analysis. To reach
this page highlight the CELL PROGRAMMER
option from the FIXED, QUANT, RATIO,
SCAN, or RATE method pages, and press
ENTER.
Alternatively, press ACCESSORIES from the
MENU page then select CELL PROG and
press ENTER.
:
Used to change the current cell position, using <>
arrow keys. This change is reflected in the Status box.
:
Selects from MANUAL / RUN&STEP / AUTO / OFF.
Cell position is only changed by use of the <> key on the
keyboard.
Allows measurement on the current cell then automatically
moves to the next cell ready for the next measurement.
Performs a measurement on each of the cells in turn.
Turns off the Cell Programmer. The instrument now behaves
as if it has a single cell only.
:
Toggles between ON and OFF with Enter.
:
Sets the number of cells to be used in the range 1 to 7.
:
Sets the number of cycles (up to 300). For example, if
set to 4 then each cell will be measured in turn
according to the application method being used four
times. MODE must be set to AUTO.
English
GD Issue 1 (08/2000)

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Helios delta

Inhaltsverzeichnis