FIXED Method Page Function Keys
* FIXED *
Pressing RUN starts a fixed measurement using the current method and
switches to the FIXED RESULTS page.
Pressing ZERO starts a zero using the current method.
Any changes to the WAVELENGTH or INTEGRATION parameters will invalidate the
current results.
If AUTOPRINTING is selected, a change to the MODE parameter will invalidate the
current results.
FIXED RESULTS Page
The layout of the page depends on the Mode and λ Select in use.
SINGLE λ
In ABS or %T modes up to 2 columns of results are displayed per
page. In CONC mode a single column of results is displayed on
each page. Results accumulate on the same page until it is full.
MULTI λ
Two columns of results are displayed per page. Results of each
Sample always start on a new page.
To move up or down pages of results use the Up/Down arrow keys.
Results are numbered sequentially from 1 to 600. Conc values are from 0.0001
to 9999.9. Any concentration value outside this range is marked "UNDER
RANGE" or "OVER RANGE".
12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
FIXED
VIEW RESULTS - Switches to the FIXED RESULTS
page.
SAVE METHOD - Brings up the SAVE page from where
the method can be saved to the Methods Library.
MENU – Switches to the MENU page.
English
GD Issue 1 (08/2000)