Herunterladen Inhalt Inhalt Diese Seite drucken

Electrical Installation; General Description - Stuart SCF1 Bedienungsanweisung

Rotator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Electrical Installation

THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the
line supply corresponds to that shown on
the rating plate.
Model
Supply requirements
SCF1
230V, 50/60Hz
The SCF1 is supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One has a U.K. 3 pin plug and the other has a 2
pin "Shuko" plug for connection to the mains.
Choose the lead appropriate for your electrical
installation and discard the other. Should neither
lead be suitable, take the lead with the U.K. plug
and replace the plug with a suitable alternative.
This involves cutting off the moulded plug,
preparing the cable and connecting to the
re-wireable plug in accordance with its
instructions.
THIS OPERATION SHOULD ONLY BE
UNDERTAKEN BY A QUALIFIED ELECTRICIAN
NOTE: Refer to the equipment's rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated.
The wires in the mains cable are coloured as
follows:
BROWN
- LIVE
BLUE
- NEUTRAL
GREEN/YELLOW - EARTH
The SCF1 is wired with a flying lead at the rear
of the instrument. This is fitted with an IEC
socket for connection of the mains lead. The
appropriate mains lead should be connected
BEFORE connection to the mains supply.
Should the mains lead need replacement, a cable
2
of 1mm
of harmonised code H05W-F connected
to an IEC320 plug should be used.
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN
Safety Advice Before Use
Power
15W

General Description

The SCF1 is a mini personal centrifuge for
centrifugation of up to 8 x 1.5ml or 2.2ml
microfuge tubes or two strips of 8 x 0.2ml tubes
at speeds of up to 6,200rpm. The rotors are
easily installed without the use of tools. The
SCF1 is very simple to use with the added
advantages of a small footprint and quiet
operation.
Do not install the SCF1 near sources of water
or gas, or near chemical storage areas.
Do not operate with wet hands.
Never move or carry the unit while it is in use
or connected to the mains supply.
Do not use damaged or deformed rotors.
Rotors with compromised structures are
subject to sudden and severe imbalances
that can damage the centrifuge and may
cause injury to the operator.
Do not leave the centrifuge in a cold room at
less than 5°C for an extended period of time.
Never centrifuge explosive or flammable
substances. The SCF1 does not provide
protection from explosions or fires
originating in the chamber.
Do not use with infectious samples. The
SCF1 is not equipped with a back-up
isolation system to contain aerosols if a tube
should leak or break during a run.
1

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis