Installation Instruction
Sight glass with seal and ventilation screw
Source container FQG66
EA01141F/00/A2/01.15
71309184
This instruction is valid for the following spare part sets:
Order number
Original spare part set
71301299
Sight glass FQG66, seal
71301303
Ventilation screw FQG66, VA, 1 piece
We recommend that the Installation Instructions be kept with the packaging at all times.
Designated use
The spare part set and Installation Instruction are used to replace a faulty unit with a functioning unit of the same type.
Use genuine parts from Endress+Hauser only.
Only original spare part sets, intended by Endress+Hauser for the measuring device, shall be used. 
The verification has to be done via W@M Device Viewer this procedure is explained below.
For some devices there is an overview of the correct spare part set inside the connection compartment cover.
If the spare part is listed there, the verification is not required.
1.
www.endress.com/deviceviewer
2.
Ser. No.:12345
Authorized personnel to carry out repairs
Authorization to carry out a repair depends on the approval of the measuring device. The table shows the respective group of persons for each.
The person who carries out the repair is responsible for safety during the work, the quality of work completed and safety of the device
after repair.
Approval of the measuring device
Without approval
With approval (for Ex. IECEx, ATEX, FM, CSA, TIIS, NEPSI)
1 = Trained customer technician, 2 = Service technician authorized by Endress+Hauser, 3 = Endress+Hauser (send measuring device back to manufacturer)
3.
=
?
4.
www.endress.com/deviceviewer
Enter the serial number (Ser. No.), search for product data and click on
"spare parts".
List of all spare parts suitable for the instrument.
Check the order number of the spare part.
Does the order number of spare part correspond to the order number in
the spare part list?
YES, the spare part can be used.
NO, the spare part may not be used.
Group of persons authorized to carry out repairs
71309184
1, 2, 3
1, 2, 3

Werbung

Inhaltsverzeichnis
loading

Inhaltszusammenfassung für Endress+Hauser FQG66

  • Seite 1 Group of persons authorized to carry out repairs Without approval 1, 2, 3 With approval (for Ex. IECEx, ATEX, FM, CSA, TIIS, NEPSI) 1, 2, 3 1 = Trained customer technician, 2 = Service technician authorized by Endress+Hauser, 3 = Endress+Hauser (send measuring device back to manufacturer)
  • Seite 2: Safety Instructions

    • Threads (e.g. of the cover for the electronics and terminal • Hot surfaces! Risk of injury! Before commencing work, allow the compartment), which have been prepared by Endress+Hauser with system and measuring device to cool down to a touchable an abrasion resistant dry lubrication should not be additionally temperature.
  • Seite 3 Sight glass with seal and ventilation screw Tool list AF17 mm AF46 mm 10 Nm 10 Nm WARNING During exchange of ventilation screw or sight glass with gasket you may be exposed to ionizing radiation. • Contact your responsible radiation safety officer for particular instructions. •...
  • Seite 4 Sight glass with seal and ventilation screw Endress + Hauser...
  • Seite 5: Schauglas Mit Dichtung Und Belüftungsschraube

    Zulassung des Messgeräts Reparaturberechtigter Personenkreis ohne Zulassung 1, 2, 3 mit Zulassung (z.B. IECEx, ATEX, FM, CSA, TIIS, NEPSI) 1, 2, 3 1 = Ausgebildete Fachkraft des Kunden, 2 = Von Endress+Hauser autorisierter Servicetechniker, 3 = Endress+Hauser (Messgerät an Hersteller zurücksenden)
  • Seite 6: Sicherheitshinweise

    • Die in der Betriebsanleitung aufgeführten Hinweise zum Transport – Mit den jeweiligen Einsatzbedingungen der Geräte vertraut. und zur Rücksendung beachten. – Bei Ex-zertifizierten Messgeräten zusätzlich im Explosionsschutz • Bei Fragen kontaktieren Sie bitte Ihre zuständige Endress+Hauser ausgebildet. Serviceorganisation. Endress + Hauser...
  • Seite 7: Austausch Der Belüftungsschraube

    Schauglas mit Dichtung und Belüftungsschraube Werkzeugliste SW17 mm SW46 mm 10 Nm 10 Nm WARNUNG Beim Austausch der Belüftungsschraube sowie einem Schauglas mit Dichtung können Sie ionisierender Strahlung ausgesetzt sein. • Den verantwortlichen Strahlenschutzbeauftragten für spezifische Anweisungen kontaktieren. • Der Strahlenschutzbehälter muss in der Stellung "AUS" gesichert sein! •...
  • Seite 8 Schauglas mit Dichtung und Belüftungsschraube Endress + Hauser...

Inhaltsverzeichnis