Removing the data output
– It is absolutely necessary to first remove the power cable from the data output.
– Carefully fold back Data Output.
– Pull out flat ribbon cable connector (7).
– Unhook clips (6) from rear wall of balance.
– Using the cover plate, close ribbon cable connector socket (7) at rear wall of balance.
Connecting the data output
– Disconnect power cable from balance.
– Using clips (6), hook on data output to openings in rear wall of
balance.
– Plug in connector (7) of flat ribbon cable. Flat ribbon cable must sag
downwards.
– Push top of data output against the balance until power supply
connector (8) is completely plugged in.
– Reconnect balance with data output to power line by plugging power
line into socket (9).
Connecting data receivers
Data receivers equipped with a current loop interface and those with an RS232C
interface can be connected at the same time to an Option 011 Data Output.
(In so doing, please note that the configuration of the data output - baud rate,
parity, transfer mode - is the same for both interfaces).
– Data receivers with a 20 mA current loop interface are connected to
socket (11).
– Data receivers with an RS232C interface are connected to socket (12).
– If switch 4 of operating mode switch (2) is in the OFF position (send on
Leerseite
transfer), a hand switch (42500) or a foot switch (46278) must be
connected to sockets (10) in order to forward transfer instructions.
2