Error codes
10.3

Dew point and wet-bulb temperature

Error code
Error
E04
Caused by temperature or
humidity error
10.4

Flow velocity

Error code
Error
E03
Flow velocity is above range
10.5

Volume flow

Error code
Error
E03
Value is above display
range.
E04
Error with flow velocity
10.6
CO
Error code
Error
E03
Temperature is above
measuring range
E01/E33
Low battery voltage or CO
sensor error
60
2
Remedial action
Please refer to paragraphs 10.1 and 10.2.
Remedial action
Take measurements within range If E03 is
still displayed, return the fan anemometer
for repair.
Remedial action
Check entered tube area.
Return the fan anemometer for repair.
Remedial action
Acclimatize the fan anemometer at room
temperature for 10 minutes. If E03 is still
displayed, return the fan anemometer for
repair.
If the battery symbol is displayed, replace
2
batteries.
Manually calibrate the CO
sensor
2

Werbung