Herunterladen Inhalt Inhalt Diese Seite drucken

Faults; Error Codes; Air Temperature; Relative Humidity - Wohler FA 410 Bedienungsanleitung

Flügelradanemometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
9

Faults

Fault
It is not possible to power
on the fan anemometer
10

Error codes

10.1

Air temperature

Error code
Error
E02
Temperature is below
measuring range
E03
Temperature is above
measuring range
E31
Temperature sensor defec-
tive
10.2

Relative humidity

Error code
Error
E04
Caused by temperature
error
E11
Calibration error
E31
Moisture sensor defective
Possible cause
On/Off key pressed too
fleetingly
Batteries not inserted cor-
rectly
Battery voltage insufficient
Remedial action
Acclimatize the fan anemometer at room
temperature for 30 minutes. If E02 is still
displayed, return the fan anemometer for
repair.
Acclimatize the fan anemometer at room
temperature for 30 minutes. If E02 is still
displayed, return the fan anemometer for
repair.
Return the fan anemometer for repair.
Remedial action
Please refer to paragraph 10.1
Return the fan anemometer for relative hu-
midity calibration
Return the fan anemometer for repair.
Remedy
Press the On/Off key slight-
ly longer
Check correct polarity
Replace batteries
Faults
EN
59

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Fa 430

Inhaltsverzeichnis