7 Troubleshooting

7 Troubleshooting
DANGER:
There is a danger of injury and possible equipment damage through incorrect power supply!
Let problems relating to an incorrect power supply be eliminated by an electrical specialist! Do not
carry out arbitrary work on the electrical components of the pyrometer!
Only eliminate such problems yourself when their causes obviously relate to incorrect power supply or contam-
ination of the lens.
Do not undertake any interventions into the pyrometer. If problems arise which do not relate to the causes
mentioned above, inform the service staff of the KLEIBER Infrared GmbH.(for contact data see (section 3.3)
at page 7)
Fault
Pyrometer does not provide any
measured values
If the pyrometer supplies inac-
curate measured values or mea-
sured values which lie outside the
range to be expected
16
Cause
Power supply faulty or interrupted
Dirty optics or condensation on
the lens
Solution
• Check the power supply
• Check plugs and connections
• Inspect cable
Clean optics (see Cleaning the
Optics (section 9.2) at page 18)
KS / KG / KGA 740 - LO