IMPORTANT:
The fixed optic cannot be adjusted. The distances mentioned must be respected absolutely in order to
ensure the small measuring field.
5.2.3 Setting the vario fiber optics
Adjust the vario fiber optics as follows:
1. If necessary switch on the pilot light.
2. Loosen the ring nut by turning it in a counter-clockwise direction.
3. Hold the screw connector of the optical fiber and turn the housing until length L has been reached,
for the correspondingly measuring distance. Mentioned distances in the data sheet are essential to be
maintained.
4. After adjusting the measuring distance, lock the ring nut by turning it in a clockwise direction.
The vario optics is adjusted to the required measuring distance.
Example (LVO 25)
a
a
in mm
b
M
in mm
c
L
in mm
a
Measuring distance
b
Measuring field diameter
c
Length
KS / KG / KGA 740 - LO
Ring nut
Housing
Optical fiber
Figure 4: Fixed optics
Optical fiber
Ring nut
Housing
Length L
Figure 5: Setting the fiber optics
80
100
120
1.6
1.8
2.0
63.2
60.0
58.1
Length L
Measuring distance
Minimum
measuring distance
Maximum measuring distance
140
160
180
200
2.2
2.4
2.6
2.8
56.2
55.4
54.6
54.0
5 Technical Description
220
240
260
280
3.1
3.4
3.7
4.0
53.6
53.1
52.7
52.4
300
4.3
52.1